Multi-Well MEA Biochips
We now offer the possibility to carry out multiple cell culture experiments in parallel on the same biochip, through our new Multi-Well MEA biochip, whose dimensions and configuration conform closely to that of the standard 96-well microplate format. The first generation of the Multi-Well Biochip comprises four wells per chip, where each well consists of 15 recording/stimulation electrodes, 1 large embedded ground electrode and 2 supplementary larger stimulation electrodes.
We welcome any enquiries and feedback on this product, as well as questions on customized biochip layouts.
|
|
The picture on the right displays the layout of the current 4-well MEA Biochip. Black: the electrodes, dark grey: Plastic insert defining the wells, light grey: electrode wires.
|
|
Blown-up version of a single well, displaying the large ground electrode, supplementary stimulation, and standard recording/stimulating electrodes.
|
|
Picture of a Multi-Well MEA Biochip inside a MEA1060 amplifier from Multi Channel Systems MCS, Reutlingen, Germany. Note the use of an MEA1060-Y1 accessory to contact the embedded ground and supplementary stimulation electrodes.
|
|
Typical recordings of spontaneous electrical activity from dissociated hippocampal neurons (from P1 rat, 31 days in vitro) on a 4-Well MEA Biochip, where each quadrant represents data from each well.
|
|
Electrical recording from one electrode located in the left-bottom quadrant of the data displayed in the picture above.
|
|
MEA60 4Wells
Number of wells: 4 Electrodes per well: 15 recording electrodes, 2 external stimulation electrodes and 1 large GND electrode
Substrate dimension: 21 mm x 21 mm x 0.7 mm Substrate material: float glass Insulation material: SU-8 epoxy, thickness 5 µm Electrode material: platinum Electrode geometry: planar Electrode dimensions: 40 µm circular electrodes Electrode spacing: 200 µm (centre to centre)
Electrode impedance at 1 kHz: 500-700 kΩ, GND electrode 450 Ω, stimulation electrodes 20-30 kΩ Noise level: 20-25 µV
Applications: Dissociated cell cultures (brain tissue, spinal cord, retina, heart muscle cells, etc.) Re-use: Several times depending on type and duration of cultures
Required accessories: MEA60 Spa-Y2 and MEA1060-Y1 or MEA1060-Y2 depending on MEA1060 amplifier from Multi Channel Systems MCS GmbH.
|
Ayanda Biosystems SA, PSE Parc Scientifique, Building C, EPFL, CH-1015 Lausanne, Switzerland
Tel/Fax: +41 21 693 8631, contact us via e-mail using this link